4- 98
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
RM
This command specifies the auto range operation for the current measurement.
Syntax
RM
chnum
,
mode
[,
rate
]
where the
rate
parameter is available for
mode
=2 or 3.
Parameters
chnum
:
Current measurement channel number. The value must be slot number
where the module has been installed. Integer expression. 1 to 8 (1 and 5
are not available for HPSMU). See
.
mode
:
Range changing operation mode. Integer expression. 1, 2 or 3.
where
current1
and
current2
are given by the following formula:
current1
=
measurement range
×
rate
/100
current2
=
measurement range
×
rate
/1000
For 200 mA range, they must be:
current1
=200 mA
×
rate
/100
current2
=100 mA
×
rate
/100
For example, if
measurement range
=10 mA and
rate
=90, these values
are as follows:
current1
= 9 mA
current2
= 0.9 mA
rate
:
Parameter used to calculate the
current
value. Numeric expression.
11 to 100. Default value is 50.
Example
Statements
OUTPUT @E5270;"RM 1,2"
OUTPUT @E5270;"RM 2,3,60"
mode
Description
1
Initial setting. If you set
mode
=1, do not set
rate
.
2
If measured data
≥
current1
, the range changes up after
measurement.
3
If measured data
≤
current2
, the range changes down
immediately, and if measured data
≥
current1
, the range
changes up after measurement.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......