Before starting the trigger in/out diagnostics, connect a BNC cable between the
Ext Trig In and Out connectors.
•
After starting the front panel key diagnostics, press any key and confirm that the
key name is displayed. Repeat this for all keys.
If all response was good, press the
Enter
key twice to stop.
result
returns 0.
If any response was bad, press the
Exit
key twice to stop.
result
returns 1.
•
After starting the high voltage LED diagnostics:
If the LED blinks, press the
Enter
key to stop.
result
returns 0.
If the LED does not blink, press the
Exit
key to stop.
result
returns 1.
•
Before starting the digital I/O diagnostics, disconnect any cable from the digital
I/O port.
•
After starting the beeper diagnostics:
If you listen 2 types of beep sounds every 1 second, press the
Enter
key to stop.
result
returns 0.
If you do not listen it, press the
Exit
key to stop.
result
returns 1.
DO
The DO command executes the E5260/E5270 internal memory programs (up to 8
programs) in the order specified.
Execution
Conditions
The specified programs have been stored by using the ST and END commands.
Syntax
DO
pnum
[,
pnum
[,
pnum
[,
pnum
[,
pnum
[,
pnum
[,
pnum
[,
pnum
]]]]]]]
Parameters
pnum
:
Internal memory program number. Numeric expression. 1 to 2000.
Example
Statements
OUTPUT @E5270;"DO 1,2,3,4,5,6,7,8"
OUTPUT @E5270;"DO 98,99"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......