12 Keysight E5053A Microwave Downconverter
Installation
Requirements for installation environment
1-
Requirements for installation environment
Install the E5053A under an environment that satisfies the following
conditions.
Operating environment
Use the E5053A under the following environmental conditions.
Temperature
10
˚
C to 40
˚
C
Temperature range for
error-correction
23
˚
C
5
˚
C (< 1
˚
C deviation from the temperature
when performing the error-correction)
Humidity
20 % to 80 % at wet bulb temperature < +28
˚
C (non-
condensation)
Altitude
0 m to 2,000 m (0 to 6,561 feet)
Vibration
0.21 G maximum, 5 Hz to 500 Hz
The above environmental requirements are not intended for the
specifications and measurement accuracy of the analyzer but for the
operating environment of the analyzer.