90
Chapter 6
Replacement Procedure
Replacing an Assembly
Replacing an Assembly
The following steps show the sequence for replacing an assembly in a E4981A
Capacitance Meter.
1. Identify the faulty group. Refer to
2. Order a replacement assembly. Refer to
Chapter 5, “Replaceable Parts.”
3. Replace the faulty assembly and determine what adjustments are necessary. Refer to
This chapter and
Chapter 7, “Post-Repair Procedures.”
4. Perform the necessary adjustments. Refer to
5. Perform the necessary performance tests. Refer to
Chapter 2, “Performance Test.”
WARNING
These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
WARNING
The opening of covers or removal of parts is likely to expose dangerous voltages.
Disconnect the instrument from its power supply.
CAUTION
Many of the assemblies in this instrument are very susceptible to damage from
ESD(electrostatic discharge). Perform the following procedures only at a static-safe
workstation and wear a grounding strap.
Tightening Torque of Screws
Please refer to the following table for the tightening torque of screws if no torque is
specified by this chapter.
Used Tool
Type
Recommended Torque
Torque Screwdriver
T6
0.34 N-m ( 3.0 lb-in)
Torque Screwdriver
T8
0.59 N-m ( 5.2 lb-in)
Torque Screwdriver
T10
1.02 N-m ( 9.0 lb-in)
Torque Screwdriver
T20
2.37 N-m (21.0 lb-in)
Box Torque Wrench
7 mm
0.98 N-m (8.7 lb-in)
Box Torque Wrench
3/8 in.
1.97 N-m (17.4 lb-in)
Box Torque Wrench
5/8 in.
1.97 N-m (17.4 lb-in)
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...