44
Chapter 4
Troubleshooting
To Troubleshoot the Instrument
Figure 4-1
Primary trouble isolation flowchart
㪼㪋㪐㪏㪈㪸㫊㪼㪈㪇㪇㪌
Power on
Power up normally?
Go to Power on Sequence
Troubleshooting
Go to Troubleshooting
Using Service Function
Go to Power on Sequence
Troubleshooting
Power on test pass?
Error message is
displayed?
Go to Function Specific
Troubleshooting*
(To Check the LCD)
Yes
No
Yes
No
Yes
Power on
Power on test
Go to Troubleshooting
Using Service Function
Power on test pass?
Yes
No
End
Performance test
Go to Performance Test
Failure Troubleshooting
Performance test pass?
Yes
No
No
* In case of display scan,
brightness and color problems.
Display and Boot-up Problems
Problems on Measurement Performance and Measurement Result
Problems on Front Panel, LCD Display,
Front USB Port, GPIB Interface, USB
Interface, LAN Port, External Trigger Input,
Handler Interface or Scanner Interface.
Power on
Go to Function Specific
Troubleshooting
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...