146
Appendix D
Error Messages
Error Messages
O
47
OPEN measurement incomplete
This error occurs when the cable correction OPEN measurement is incomplete.
-321
Out of memory
The E4981A has insufficient memory to perform the requested operation.
P
1080
Power failed
Power unit hardware failure is detected.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
-108
Parameter not allowed
The number of parameters is larger than required by the command. For example, although
the
:CREJ:LIM
command requires one parameter such as “
:CREJ:LIM 3
”, two
parameters are added to give “
:CREJ:LIM 0,3
”.
-112
Program mnemonic too long
The length of the header exceeds 12 characters. (Refer to IEEE488.2,7.6.1.4.1.)
Q
-350
Queue overflow
The queue contains a certain code other than the code that caused this error. This indicates
that an error has occurred due to insufficient space in the queue but has not been recorded.
-400
Query error
A comprehensive Query error has occurred for which the E4981A cannot detect further
details. This code simply indicates the occurrence of a Query error that is defined in
IEEE488.2,11.5.1.1.7 and 6.3.
-410
Query INTERRPUTED
This indicates the status that causes an “INTERRUPTED” Query error. (Refer to
IEEE488.1,6.3.2.3.) This error occurs, for example, when data byte (DAB) or GET is
received after Query but before the response has been completely sent.
-420
Query UNTERMINATED
This indicates the status that causes an “UNTERMINATED” Query error. (Refer to
IEEE488.2,6.3.2.) This error occurs, for example, when the E4981A is specified as a talker
and an incomplete program message is received.
-430
Query DEADLOCKED
This indicates the status that causes a “DEADLOCKED” Query error. (Refer to
IEEE488.2,6.3.1.7.) This error occurs, for example, when both input and output buffers
become full and the E4981A cannot continue processing.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...