144
Appendix D
Error Messages
Error Messages
F
1070
Fan failed
Cooling fan hardware failure is detected.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
G
-105
GET not allowed
A group execution trigger (GET) has been received in a program message. (Refer to
IEEE488.2,7.7.)
H
-114
Header suffix out of range
The header suffix is out of range.
I
-101
Invalid character
Invalid characters have been found in the program message string. For example, in a
correct program message “
:CALC1:FORM CP
”, an ampersand (&) is inserted by mistake
to give “
:CALC1:FORM&CP
”.
-103
Invalid separator
The parser (syntax analysis program) expects a separator, but a character other than a
separator has been sent. For example, although the correct way is to use “;” to separate two
sent program messages such as “
:CALC1:FORM CP;*OPC?
”, the semicolon (;) needed to
separate the program messages is missing to give “
:CALC1:FORM CP *OPC?
”.
-121
Invalid character in number
An invalid character for the data type of the syntax analysis target has been received. For
example, alphabetical characters exist in a decimal value or “9” exists in octal data.
-131
Invalid suffix
The suffix does not meet the syntax defined in IEEE488.2,7.7.3.2 or is inappropriate for
the E4981A.
-141
Invalid character data
A character data element has been recieved where the E4981A does not accepts any
character data element.
-151
Invalid string data
Character string data are expected, but the string data received are invalid for some reason.
(Refer to IEEE488.2,7.7.5.2.) For example, the END message is received before the end
quotation mark character appears.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...