142
Appendix D
Error Messages
Error Messages
Error Messages
An error message is displayed in the instrument status display area in the lower-left part of
the screen. Pressing the cursor arrow key on the front panel or executing the :DISP:CCL
command clears the error message. Moreover, about a specific error message, when a
sweep is started again, the display of an error message may disappear. Errors caused by
operation of a front panel key simply appear on the display; with a few exceptions, these
are not stored in the error queue.
The log of an error message, a maximum of 100 pieces are recorded, and it can go back and
check from the oldest error. Executing the :SYST:ERR? command checks error logs. This
operation can be performed only by "SCPI" command. It cannot be operated from the front
panel.
An error with a positive error number is one uniquely defined for this instrument. On the
other hand, an error with a negative error number is basically one defined for common
GPIB devices in IEEE488.2.
1 - 100
21
1 MHz opt. not installed
This error occurs when cable correction commands are sent to E4981A with Option 002,
via GPIB/LAN/USB. This error is not generated with front panel operation.
A
1103
A1 EEPROM write error
An error is generated while writing data to A1 EEPROM.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
B
-168
Block data not allowed
A block data element has been received where the E4981A does not accept any block data
element.
C
41
Correction Measurement Aborted
This error occurs when the correction data measurement is aborted.
1200
CPU bd FLASH ROM write error
An error is generated while writing data to FLASH.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...