56
Chapter 4
Preparation for Accurate Measurement (Executing Correction)
Using the Multi-correction Function
Using the Multi-correction Function
Turning ON/OFF the multi-correction function
To turn ON/OFF the multi-correction function, use the following command.
•
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
Selecting a channel
To select a channel for the multi correction function, use the following command.
•
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
Measuring correction data
Selecting the definition method of the LOAD correction standard
To select whether to define the LOAD correction standard value (LOAD correction
reference value) for each channel individually or for all channels commonly, use the
following command.
•
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] on page 240
Measuring correction data
The method to measure the OPEN/SHORT/LOAD correction data for multi-correction is
the same as that for basic correction data except that you need to select the proper channel
before measurement. For more details, refer to
“Measuring correction data” on page 52
Reading/writing correction data
The method used to read/write the OPEN/SHORT/LOAD correction data for
multi-correction is the same as that for basic correction data except that you need to select
the proper channel before measurement. For more details, refer to
correction data (saving/recalling correction conditions)” on page 53
Sample Program
to view an example of multi-correction.