Chapter 4
55
Preparation for Accurate Measurement (Executing Correction)
Executing Offset Correction
4. Pr
ep
ar
at
ion
fo
r
Accu
ra
te
M
eas
ur
em
en
t
(Executing
C
orrection)
Executing Offset Correction
Turning ON/OFF the correction function
When you turn ON the offset correction, assuming that the measured value before
correction is
and the offset correction data is
, the measured value is
compensated as
.
To turn ON/OFF the offset correction, use the following command.
•
[:SENSe]:CORRection:OFFSet[:STATe] on page 243
You cannot turn ON/OFF the primary parameter and secondary parameter separately.
However, if you set the correction value to 0, the state is actually the same as OFF even if
the offset correction is ON. Therefore, in practice you can realize separate ON/OFF states
by setting the correction value for either parameter to 0.
Setting up the correction data
To set up the offset correction data, use the following command.
•
[:SENSe]:CORRection:OFFSet:DATA on page 242
The entered value is set as the offset correction data for the measurement frequency at the
time of the entry.
Meas
Offset
Meas Offset
–