:CALCulate1:COMParator:COUNt:DATA?
Syntax
:CALCulate1:COMParator:COUNt:DATA?
Description
Reads out each count value of BIN1 to BIN9, OUT_OF_BINS, and AUX_BIN of the BIN
counter function.
Reads out all BIN count values regardless of the ON/OFF state of each BIN (set with the
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe
command). (Query only)
Parameters
Query response
{numeric 1},…,{numeric 11}<newline><^END>
Related commands
:CALCulate1:COMParator:COUNt:OVLD? on page 179
:CALCulate1:COMParator:COUNt[:STATe] on page 180
:CALCulate1:COMParator:COUNt:CLEar on page 176
:CALCulate1:COMParator:COUNt:MULTiple:DATA? on page 178
:CALCulate1:COMParator[:STATe] on page 187
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe on page 183
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
{numeric 1}
The count value of BIN1.
{numeric 2}
The count value of BIN2.
{numeric 3}
The count value of BIN3.
{numeric 4}
The count value of BIN4.
{numeric 5}
The count value of BIN5.
{numeric 6}
The count value of BIN6.
{numeric 7}
The count value of BIN7.
{numeric 8}
The count value of BIN8.
{numeric 9}
The count value of BIN9.
{numeric 10}
The count value of OUT_OF_BINS.
{numeric 11}
The count value of AUX_BIN.