VOLTage Subsystem
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<voltage>|MINimum|MAXimum|DEFault
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum | MAXimum]
Sets the output voltage in volts.The query returns a number of the form +#.########E+##.
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement] <numeric
value>|DEFault
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]? [DEFault]
Sets the step size for voltage programming with the VOLTage UP and VOLTage DOWN commands. The
query returns a number of the form +#.########E+##.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <voltage>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the triggered voltage level in amps. The query returns a number of the form +#.########E+##.
[SOURce:]VOLTage:PROTection:CLEar
Clears an overvoltage protection event.
[SOURce:]VOLTage:PROTection:STATe ON|1|OFF|0
[SOURce:]VOLTage:PROTection:STATe?
Enables or disables overvoltage protection, which causes the instrument to go into a protected state
when the power supply goes above the protection level specified by the VOLTage:PROTection[:LEVel].
The query returns 1 (ON) or 0 (OFF).
[SOURce:]VOLTage:PROTection:TRIPped?
Indicates whether an overvoltage protection occurred (1) or not (0). This is reset to 0 by
VOLTage:PROTection:CLEar.
[SOURce:]VOLTage:PROTection[:LEVel] <
voltage
>|MINimum | MAXimum
[SOURce:]VOLTage:PROTection[:LEVel]? [MINimum|MAXimum]
Sets the level at which overvoltage protection trips, in volts. The query returns a number of the form
+#.########E+##.