30
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify time base accuracy
This test verifies the accuracy of the time base. In this test you will measure the
absolute error of the time base oscillator and compare the results to the
specification.
1
Set up the signal generator.
a
Set the output to 10 MHz, approximately 1 V
pp
sine wave.
2
Connect the output of the signal generator to oscilloscope channel 1 using the
BNC cable. Also, connect a 50 ohm feedthrough termination between the
channel 1 input and the BNC cable.
3
Set up the oscilloscope:
a
Press
[AutoScale]
.
b
Set the oscilloscope Channel 1 vertical sensitivity to 200 mv/div.
c
Set the oscilloscope horizontal sweep speed control to 5 ns/div.
d
Adjust the intensity to get a sharp, clear trace.
e
Adjust the oscilloscope’s trigger level so that the rising edge of the
waveform at the center of the screen is located where the center horizontal
and vertical grid lines cross (center screen).
f
Ensure the horizontal position control is set to 0.0 seconds.
Table 8
Equipment Required to Verify Time Base Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
100 kHz - 100 MHz, 0.01 Hz frequency
resolution,
jitter: < 2ps
Keysight N5181A
Cable
BNC, 3 feet
Keysight 10503A
50 Ohm Feedthrough
Termination
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...