26
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify bandwidth (-3 dB)
This test checks the bandwidth (-3 dB) of the oscilloscope. In this test you will use
a signal generator and a power meter.
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the input of the power
splitter input.
b
Connect the power sensor to one output of the power splitter.
c
Use an N-to-BNC adapter to connect the other splitter output to the
channel 1 input using a 50 ohm feedthrough terminator at the oscilloscope
input BNC.
Table 6
Bandwidth (-3 dB) Test Limits
Models
Test Limits
100 MHz Models
All channels (-3 dB), dc to 100 MHz
70 MHz Models
All channels (-3 dB), dc to 70 MHz
50 MHz Models
All channels (-3 dB), dc to 50 MHz
Table 7
Equipment Required to Verify Bandwidth (-3 dB)
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
100 kHz - 100 MHz at 200 mVrms
Keysight N5181A
Power Meter
1 MHz - 100 MHz ±3% accuracy
Keysight N1914A
Power Sensor
1 MHz - 100 MHz ±3% accuracy
Keysight E9304A or N8482A
Power Splitter
outputs differ by < 0.15 dB
Keysight 11667A
Cable
Type N (m) 24 inch
Keysight 11500B
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082 or
Pomona 3288 with Pomona
3533
50 Ohm Feedthrough
Termination
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...