CAN/LIN Triggering and Serial Decode
23
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
333
• Aliased bus values (undersampled or indeterminate) are drawn in pink.
• Unknown bus values (undefined or error conditions) are drawn in red with a "?"
label.
• Flagged error frames are drawn in red with an "ERR" label.
CAN Totalizer
The CAN totalizer provides a direct measure of bus quality and efficiency. The CAN
totalizer measures total CAN frames, flagged error frames, overload frames, and
bus utilization.
The totalizer is always running (counting frames and calculating percentages) and
is displayed whenever CAN decode is displayed. The totalizer counts even when
the oscilloscope is stopped (not acquiring data). Pressing the
[Run/Stop]
key does
not affect the totalizer. When an overflow condition occurs, the counter displays
OVERFLOW
. The counters can be reset to zero by pressing the
Reset CAN Counters
softkey.
Types of Frames
• Active error frames are CAN frames in which a CAN node recognizes an error
condition during a data or remote frame and issues an active error flag.
• A partial frame occurs when the oscilloscope detects any error condition during
a frame that is not followed by an active error flag. Partial frames are not
counted.
Counters
• The FRAMES counter gives the total number of completed remote, data,
overload, and active error frames.
• The OVLD counter gives the total number of completed overload frames and
their percentage of the total number of frames.
• The ERR counter gives the total number of completed active error frames and
their percentage of the total number of frames.
• The UTIL (bus load) indicator measures the percentage of time the bus is active.
The calculation is done on 330 ms periods, approximately every 400 ms.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...