Utility Settings
20
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
287
To perform user calibration
Perform user-calibration:
• Every two years or after 4000 hours of operation.
• If the ambient temperature is >10° C from the calibration temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self-alignment routine to optimize the signal path in
the oscilloscope. The routine uses internally generated signals to optimize circuits
that affect channel sensitivity, offset, and trigger parameters.
Performing User Cal will invalidate your Certificate of Calibration. If NIST (National
Institute of Standards and Technology) traceability is required, perform the
"Performance Verification" procedure in the
Keysight InfiniiVision
2000/3000 X-Series Oscilloscopes Service Guide
using traceable sources.
To perform user calibration:
1
Disconnect all inputs from the front and rear panels, including the digital
channels cable on an MSO, and allow the oscilloscope to warm up before
performing this procedure.
2
Press the rear-panel CAL button to disable calibration protection.
3
Connect short (12 inch maximum) equal length cables to each analog channel's
BNC connector on the front of the oscilloscope. You will need two equal-length
cables for a 2-channel oscilloscope or four equal-length cables for a 4-channel
oscilloscope.
Use 50W RG58AU or equivalent BNC cables when performing User Cal.
For a 2-channel oscilloscope, connect a BNC tee to the equal length cables.
Then connect a BNC(f)-to-BNC(f) (also called a barrel connector) to the tee as
shown below.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...