278
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
20
Utility Settings
If Automatic is not enabled, the oscilloscope's LAN configuration must be set
up manually using the
Addresses
and
Host name
softkeys
5
Configure the oscilloscope's LAN interface:
a
Press the
Addresses
softkey.
b
Use the
Modify
softkey (and the other softkeys and the Entry knob) to enter
the IP Address, Subnet Mask, Gateway IP, and DNS IP values. When you are
done, go back up in the menu hierarchy.
c
Press the
Host name
softkey. Use the softkeys and the Entry knob to enter the
Host name. When you are done, go back up in the menu hierarchy.
d
Press the
Apply
softkey.
6
Connect the oscilloscope to the local area network (LAN) by inserting the LAN
cable into the "LAN" port on the rear panel of the oscilloscope.
Stand-alone (Point-to-Point) Connection to a PC
The following procedure describes how to establish a point-to-point (stand alone)
connection to the oscilloscope. This is useful if you want to control the
oscilloscope using a laptop computer or a stand-alone computer.
1
Press
[Utility] > I/O
.
2
Press the
LAN Settings
softkey.
3
Press the
Config
softkey; then, turn the Entry knob to select
Automatic
, and press
the softkey again to enable it.
If your network supports DHCP or AutoIP, enabling
Automatic
lets the
oscilloscope use those services to get its LAN configuration settings
4
Connect your PC to the oscilloscope using a cross-over LAN cable such as
Keysight part number 5061-0701, available on the web at
5
Cycle power on the oscilloscope. Wait until the LAN connection is configured:
•
Press
[Utility] > I/O
and wait until the LAN status shows "configured".
This may take a few minutes.
Now, the instrument is connected, and the instrument's web interface or remote
control via LAN may be used.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...