Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
139
You can push the Trigger Level knob to set the level to the waveform's 50% value.
If AC coupling is used, pushing the Trigger Level knob sets the trigger level to
about 0 V.
The position of the trigger level for the analog channel is indicated by the trigger
level icon (if the analog channel is on) at the far left side of the display. The
value of the analog channel trigger level is displayed in the upper-right corner of
the display.
The trigger level for a selected digital channel is set using the threshold menu in
the Digital Channel Menu. Press the
[Digital]
key on the front panel, then press the
Thresholds
softkey to set the threshold level (TTL, CMOS, ECL, or user defined) for
the selected digital channel group. The threshold value is displayed in the
upper-right corner of the display.
The line trigger level is not adjustable. This trigger is synchronized with the power
line supplied to the oscilloscope.
Forcing a Trigger
The
[Force Trigger]
key causes a trigger (on anything) and displays the acquisition.
This key is useful in the Normal trigger mode where acquisitions are made only
when the trigger condition is met. In this mode, if no triggers are occurring (that is,
the "Trig'd?" indicator is displayed), you can press
[Force Trigger]
to force a trigger
and see what the input signals look like.
In the Auto trigger mode, when the trigger condition is not met, triggers are forced
and the "Auto?" indicator is displayed.
T
NOTE
You can also change the trigger level of all channels by pressing
[Analyze] > Features
and
then selecting
Trigger Levels
.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...