Preparing to Take Measurements
26
DDR2(+LP) Compliance Testing Methods of Implementation
Calibrating the Oscilloscope
If you haven’t already calibrated the oscilloscope and probe, see
NOTE
If the ambient temperature changes more than 5 degrees Celsius from the calibration temperature,
internal calibration should be performed again. The delta between the calibration temperature and
the present operating temperature is shown in the Utilities>Calibration menu.
NOTE
If you switch cables between channels or other oscilloscopes, it is necessary to perform cable and
probe calibration again. Keysight recommends that, once calibration is performed, you label the
cables with the channel on which they were calibrated.