6-16
Keysight B2900 User’s Guide, Edition 5
Function Details
Resistance Measurement
Resistance Measurement
B2900 supports resistance measurement. If the measurement parameter is set to
resistance OHMS (R), the source/measure unit (SMU) automatically sets the current
source and voltage measurement operation to perform resistance measurement.
For performing accurate measurement, B2900 provides the compensation function.
NOTE
To Enable Resistance Measurement
The resistance measurement operation must be specified. See
“To Enable Resistance Measurement” on page 5-22
In the initial setting, the resistance measurement operation is set to OFF.
Resistance Compensation
Resistance compensation (R Compen) is effective for performing low resistance
measurements accurately. If R Compen is set to ON, the channel performs
measurement twice and returns the compensated measurement result given by the
following formula. This technique is effective for reducing the thermal EMF.
R
compen
=(V
2
-
V
1
)/(I
2
-
I
1
)
where, V
1
is the measurement results at the 0 A source condition, and I
1
is the
measurement results at the 0 V source condition.
To enable resistance compensation, press the Config > Measure > R Compen > ON
function key on a 1-channel model, or the Config > Measure > R Compen > ALL or
Ch 1 or Ch2 > ON function key on a 2-channel model.
To disable resistance compensation, press the Config > Measure > R Compen >
OFF function key on a 1-channel model, or the Config > Measure > R Compen >
ALL or Ch 1 or Ch2 > OFF function key on a 2-channel model.
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......