Keysight B2900 User’s Guide, Edition 5
4-17
Front Panel Reference
Display and Assist Keys
Trigger parameters
The B2900 supports the following trigger types for triggering source output and
measurement. They are effective for setting triggers easily.
The trigger type can be selected by using the assist keys displayed when the field
pointer is in the EDIT (green) status on the Trigger field. See
for the
trigger types and the setup parameters.
AUTO
Automatic trigger type
SYNC
Synchronize trigger type
TIMER
Timer trigger type
MANUAL
Manual trigger type
The following setup parameters are available for setting the source output trigger
(transient action) by using the Source column, and the measurement trigger (acquire
action) by using the Measure column.
Count
Trigger count (number of triggers), 1 to 100000 or INF.
(infinity). This value is automatically set for the AUTO trigger
type. For the other trigger type, set the number of triggers
required for each source output and measurement properly. For
example, set Source Count = Measure Count = 10 for the sweep
measurement with 10 steps.
Delay
Sets the trigger delay time.
Period
Sets the trigger period.
Trigger
Selects the trigger source by using the assist key AUTO, BUS,
TIMER, INT
m
(only for 2-channel model), LAN, or EXT
n
,
where
m
is the integer 1 or 2, and
n
is an integer from between 1
and 14. See
for the trigger sources.
When the field pointer is in the EDIT (green) status on the Delay or Period field, the
assist keys are changed to the unit keys as shown below.
s, ms, s, and ks
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......