Installation
Inspection and Installation
To Enable System Controller
To allow this instrument to control external GPIB devices, perform the following
procedure. After the procedure, you cannot control the instrument from an external
computer.
1. If the EasyEXPERT software is running, terminate it as shown below:
a. Select the menu function
File
>
Exit
on the EasyEXPERT main screen.
b. Click [x] at the upper right corner of the Start EasyEXPERT button.
2. Select
All Programs
>
Keysight IO Libraries Suite
>
Keysight Connection
Expert
from the Start menu. Keysight Connection Expert window appears.
3. At the
Instrument I/O on this PC
area, select GPIB0, and click the Change
Properties... button. Keysight 82350 PCI GPIB Interface - GPIB0 window
appears.
4. Set the GPIB Address value to 21 that is the typical address number for the
system controller.
5. Check the
System Controller
box.
6. Remove the check from the
Auto-discover instruments connected to this
interface
box.
This disables the automatic device detection. So the Connection Expert cannot
detect the instruments connected to the GPIB interface by itself.
7. Click the OK button on Keysight 82350 PCI GPIB Interface - GPIB0 window.
8. On the Reboot Required dialog box, click the Reboot Now button, and reboot
the mainframe.
NOTE
To Disable System Controller
To disable the system controller, change the steps 4 and 5 for setting the GPIB
Address value to
not
21 and removing the check from the
System Controller
box,
and perform the procedure to the step 8.