5-22
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Dual HCSMU Adapter
To Connect 16493S-021
The 16493S-021 adapter is used with an accessory, 16493S-010/011 HCSMU adapter,
N1258A module selector, or N1259A/N1265A test fixture. The adapter should be installed
near the accessory.
Table 5-14
To Connect 16493S-021 Input
Table 5-15
To Connect 16493S-021 Output
CAUTION
If DHCSMU is used with the N1258A or N1259A-300 module selector, the maximum
current must be ± 30 A to prevent the module selector from performance degradation and
failure.
B1505A terminals
Required cables Input
terminals
HCSMU
used for
master
Force
16493S HCSMU cable, 3 m or 1.5 m
Master
Force
Sense
Sense
HCSMU
used for
slave
Force
16493S HCSMU cable, 3 m or 1.5 m
Slave
Force
Sense
Sense
Output terminals
1
1. This is the DHCSMU channel. The connectors are compatible with HCSMU.
Required cables/Description
Connect to
Output
Force
dedicated cable, 300 mm, furnished with 16493S-021
To connect N1265A, adapters N1254A-103 for Force and
N1254A-517 for Sense are required.
HCSMU input
connectors of
16493S-010/011,
N1258A, N1259A,
or N1265A
Sense
16493S
Opt 021
Dual HCSMU Combination Adapter
Force
Sense
High Sense
Low Sense
HCSMU
Circuit
Common
High Force
Low Force
±
40 V Max
Output
Sense
Force
Sense
Force
Master
Slave
From HCSMU
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......