5-8
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
Connection Overview
Table 5-4
To Connect N1267A Fast Switch
B1505A terminals
Required cables
N1267A terminals
MCSMU
Force
16494A Triaxial cable, 3 m or 1.5 m, 1 ea.
MCSMU
Force
Sense
16494A Triaxial cable, 3 m or 1.5 m, 1 ea.
Sense
HCSMU
Force
16493S HCSMU cable, 3 m or 1.5 m, 1 ea. for HCSMU
HCSMU
Force
Sense
Sense
HVSMU
Force
16493T HVSMU cable, 3 m or 1.5 m, 1 ea.
HVSMU
GNDU
16493L GNDU cable, 3 m or 1.5 m, 1 ea.
GNDU
N1267A terminals
Required cables/Description
Connect to
Low
Force
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
DUT low terminal
(ex: source)
High
Force
Coaxial cable with HV(plug) connector. N1254A-506 can be used.
Use coaxial cable to minimize affect of inductance.
DUT high terminal
(ex: drain)
Shielding box
B 1505A
DUT interface
N1267A
Interlock circuit
16493T
16493L
Low Force
High Force
*2
*1
*3
*1: 1252-1419
*2: N1254A-506 or other cable, 1 ea.
*3: N1254A-503 or other cable, 2 ea.
16493S
16494A
16493J
N1266A
*5
*5: 16494A, 4 ea., 16493L, 16493T, and 16493G
to configure HVMCU.
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......