Conduct M9202A Operational check (optional)

Conduct M9202A Operational check (optional)

Requirements for Verification

The M9202A is verified by using it to trigger on and visualize a signal from a Function Generator. The trigger
must be stable and the signal frequency and amplitude must correspond to that set on the generator.

Required Hardware

To verify that the module works requires an external signal source which will be sent to both the input and the
trigger. Almost any sine wave or function generator capable of generating a signal with a Peak-Peak Amplitude in
the range 0.1-2 V at a frequency in the range 100-500 MHz can be used. The rest of this description will assume
that an Agilent N5181A MXG RF Analog Signal Generator is available for this purpose.

Hardware

Description

Agilent N5181A

MXG RF Analog Signal Generator

1x BNC - BNC cable

50 Ω Coaxial BNC(m) - BNC(m) cable (min 20 cm)

1x Agilent 11667L Splitter

DC to 2 GHz Power Splitter with BNC connector

1x BNC - SMA cable

50 Ω Coaxial BNC(m) to SMA(m) cable (100 cm)

1x Agilent 1250-1476 Adaptor

Type N(m) to BNC(f) adaptor

1x BNC - MMCX cable

50 Ω Coaxial (100 cm)

Operational Verification Procedure

Do not exceed the maximum power level to the INPUT connector (0.5 W).

1. Launch the soft front panel (SFP). The calibration must successfully end as noted above.
2. Configure the RF Generator to produce a Sine signal with a Frequency of 100 MHz, an Amplitude of 560

mV pp (+8.0 dBm), and an Offset of 0 V.

3. Connect the RF GENERATOR output to the Splitter with the Adaptor and the BNC-BNC cable. Using the

BNC-SMA cable, connect one Splitter output to INPUT 1. Use the BNC-MMCX cable to connect the other
Splitter output to the TRG1 input. Turn on the RF GENERATOR output.

4. Configure the SFP for the Oscilloscope Mode.

5. Select the 'Connect data points' option from the 'View' menu.
6. Configure the Input for:

l

Trigger Level 1 –

0.0 V

l

Input Voltage Offset –

0.0

l

Trigger Delay –

0.0

l

Trigger Source –

M9202A EXT1

l

Timebase –

5 ns/Div

l

Trigger Coupling –

DC

l

Segments –

1

l

Trigger Slope –

Positive

l

Memory –

100

7. Select the NORMAL Acquisition mode and the Waveform Display should be as shown below.
8. Verify that:

o

the waveform shown is stable from one acquisition to the next,

o

the period of the signal is 2 divisions

o

the pk-pk amplitude is 8 ± 0.6 divisions

16

M9202A PXI Express 12-bit Wideband IF Digitizer Startup Guide

Summary of Contents for Agilent Technologies PXI Express M9202A

Page 1: ...bit Wideband IF Digitizer M9202A Startup Guide NOTICE This document contains references to Agilent Technologies Agilent s former Test and Measurement business has become Keysight Technologies For more...

Page 2: ...or contract clause Use duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will recei...

Page 3: ...the voltage being measured The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired Do not exceed th...

Page 4: ...nt means caution risk of danger You should refer to the operating instructions located in the user documen tation in all cases where the symbol is marked on the instrument This symbol indicates the ti...

Page 5: ...anel Connectors 13 Front Panel LED 13 Step 5 Verify Operation of the M9202A Module 14 Conducting a Self Test 15 Conduct M9202A Operational check optional 16 Requirements for Verification 16 Required H...

Page 6: ...Documentation Map 6 M9202A PXI Express 12 bit Wideband IF Digitizer Startup Guide...

Page 7: ...elect Technical Support Manuals l Help system for the Soft Front Panel l Help systems for the device drivers IVI C and IVI COM and LabVIEW G l Specifications Data Sheet Follow the Startup Sequence Thi...

Page 8: ...spect for damage After unpacking the module carefully inspect it for any shipping damage Report any damage to the shipping agent immediately as such damage is not covered by the warranty Warranty info...

Page 9: ...instrument or prevent it from moving in the shipping container Styrene pellets can also cause equipment damage by generating static electricity or by lodging in fan motors If any correspondence is req...

Page 10: ...r operation than is required for installation The amount of space listed above is required for installation 2 Note NET Framework Runtime Components are installed by default with Windows Vista Therefor...

Page 11: ...llowing l Agilent IO Library Suite IOLS which includes the Agilent Connection Expert This software must be installed first Version 16 2 or newer of the Agilent IO Libraries Suite is required l Soft Fr...

Page 12: ...e triangle symbol 5 Install the module into the appropriate slot of the chassis by placing the module card edges into the front mod ule guides top and bottom Ensure that the ejector handle is pushed d...

Page 13: ...t This input is AC cou pled and 50 terminated The recommended signal level is 1 V pk pk 4 dBm The clock signal frequency should be half of the desired sampling frequency TRG OUT This MMCX female conne...

Page 14: ...202 module is recognized 3 Select the M9202A module and check that the correct Model Name appears in the Instrument Properties 4 Select the Installed Software tab and review the configuration data as...

Page 15: ...e with the processor controlling it When the SFP is launched it will initiate a self calibration which takes a little over 1 minute Verify that no error is shown in the lower left hand log area The im...

Page 16: ...or 1x BNC MMCX cable 50 Coaxial 100 cm Operational Verification Procedure Do not exceed the maximum power level to the INPUT connector 0 5 W 1 Launch the soft front panel SFP The calibration must succ...

Page 17: ...cilloscope 3 If the problem occurs during Operational Verification disconnect the input signal and note the status of the LED on the front panel it should be GREEN Reconnect the signal and select the...

Page 18: ...ers Manufacturers Group PCIe and PCI EXPRESS are registered trademarks and or service marks of PC SIG Microsoft Windows VisualStudio VisualC VisualC and VisualBasic are either registered trademark or...

Reviews: