Standard Event Status Group
These registers are programmed by Common commands. The group consists of an Event and Enable register. The
Standard Event event register latches events relating to communication status. It is a read-only register that is
cleared when read. The Standard Event enable register functions similarly to the enable registers of the Operation
and Questionable status groups. See
for a description of each register.
The event status register bits are set when certain events occur during instrument operation.
The register is controlled by the IEEE 488 commands *ESE, *ESE?, and *ESR?.
Bit Value Bit Name
Description
0
1
Operation Complete
(OPC)
All commands before and including *OPC have completed.
1
2
(not used)
0 is returned
2
4
Query Error (QYE)
The instrument tried to read an empty output buffer, a command was received before a previous query was
read, or the input and output buffers are full.
3
8
Device Dependent
Error (DDE)
A device-specific error occurred.
4
16 Execution Error (EXE) An execution error occurred.
. A valid SCPI command may not be executed correctly
depending on the instrument's conditions.
5
32 Command Error
(CME)
A command syntax error occurred.
6
64 Reserved
0 is returned
7
128 Power ON (PON)
Power has been cycled since the last time the event register was read or cleared.
Keysight AC6800 Series Operating and Service Guide
99