![Keysight Technologies 86120B User Manual Download Page 239](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930239.webp)
237
Specifications and Regulatory Information
Specifications
Selectivity
Two lines input separated by
≥
100 GHz
(characteristic)
25 dB (characteristic)
Two lines input separated by
≥
30 GHz
(characteristic)
10 dB (characteristic)
Input Power
Maximum displayed level (sum of all lines)
+10 dBm
Maximum safe input level (sum of all lines)
+18 dBm
Maximum Number of Laser Lines Input
100
Input Return Loss
With flat contacting connectors
35 dB
With angled contacting connectors (Option 022)
50 dB
Measurement Cycle Time
Normal update mode
(characteristic)
1.0 s (1 measurement-per-second)
Fast update mode
(characteristic)
0.33 s (3 measurements-per-second)
Measurement Applications
Signal-to-Noise, channel spacing 200 GHz (characteristic)
>35 dB (0.1 nm noise bandwidth), lines above
–25 dBm
Signal-to-noise with Averaging, modulated lasers, 100 averages,
channel spacing 200 GHz (characteristic)
>35 dB (0.1 nm noise bandwidth) lines above -
25 dBm
Coherence Length
(characteristic)
Fabry-Perot lasers, 1 to 200 mm coherence
length, accuracy
±
5%, 0.75s cycle time
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......