Programming Commands
CALCulate3 Subsystem
SNR:REFerence:FREQuency
Enters a frequency that can be used for the noise measurement reference in signal-
to-noise calculations.
Syntax
:CALCulate3:SNR:REFerence:FREQuency{?| {<real> | MINimum | MAXimum}}
<real>
is a frequency value that is within the following limits:
Constant
Description
MINimum
181.6924 THz
MAXimum
428.2750 THz
Attribute Summary
Preset State: unaffected by
*RST State: 193.4145 THz (1550.0 nm in a vacuum).
SCPI Compliance: instrument specific
Description
After entering this value, use the SNR:AUTO command to configure the instrument
to use this value in subsequent signal-to-noise calculations.
The default units for the
<real>
parameter are Hz.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......