85053B 3-13
Use, Maintenance, and Care of the Devices
Gaging Connectors
Figure 3-3
Zeroing the Connector Gage Using the Short
4.
Assemble the airline and center conductor (refer to
a.
Remove the center conductor from its plastic case. Make sure you
select the correct center conductor for the airline you are
for
illustrations of both center conductors.
b.
Remove the protective end cap from the female end (the end
without the connecting nut) of the outer conductor.
c.
Leave the protective end cap on the male end of the airline to
prevent the center conductor from falling out of the outer
conductor.
d.
Insert the center conductor into the outer conductor so that the
female end of the center conductor is toward the female end of the
outer conductor (the end without the connector nut). Refer to
Wear gloves while performing the following steps. You will be touching the
exposed center conductor of the airline. Do not transfer oil or dirt from
your fingers to the center conductor. See
.
Summary of Contents for 85053B
Page 1: ...User s and Service Guide 3 5 mm Verification Kit Keysight 85053B ...
Page 10: ...1 6 85053B General Information Preventive Maintenance 1 ...
Page 38: ...3 22 85053B Use Maintenance and Care of the Devices Handling and Storage 3 ...
Page 42: ...5 2 85053B Troubleshooting Troubleshooting Process 5 Figure 5 1 Troubleshooting Flowchart ...
Page 50: ...6 6 85053B Replaceable Parts Replaceable Parts 6 Figure 6 1 Component Identification Sheet ...
Page 56: ...85053 90028 Index Index 6 ...