2
Service
150
Keysight 53210A/53220A/53230A Assembly Level Service Guide
– Consult the “Error Messages” section on
for what to do after
receiving a specific calibration error message.
To view the calibration count
1
Press the
Utility
key and then press the
Cal/AutoCal
softkey.
2
The display shows:
–
Cal Count
shows the total number of times: (1) a successful calibration step
has been completed; (2) calibration data is downloaded; (3) the Cal String
or Secure Code is updated; (4) the counter firmware is updated. It is not
operator-modifiable and rolls over at (2
32
-1). Since the unit is calibrated
several times during manufacture, the Cal Count will not be zero when
shipped.