Page 1: ...Keysight 16192A Parallel Electrode SMD Test Fixture Operation and Service Manual Agilent 16192A Specs Provided by www AAATesters com...

Page 2: ......

Page 3: ...Agilent 16192A Parallel Electrode SMD Test Fixture Operation and Service Manual Agilent Part No 16192 90040 Printed in JAPAN May 2003 Seventh Edition...

Page 4: ...ght All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the prior written consent of the Agilent Technologies Agilent Technologies...

Page 5: ...cause the date to change The manual part number changes when extensive technical changes are incorporated December 1993 First Edition June 1996 Second Edition October 1996 Third Edition part number 1...

Page 6: ...ages may exist even with the power cable removed To avoid injuries always disconnect power and discharge circuits before touching them DO NOT Service Or Adjust Alone Do not attempt internal service or...

Page 7: ...ution s calibration facility or to the calibration facilities of other International Standards Organization members Warranty This Agilent Technologies instrument product is warranted against defects i...

Page 8: ...ms the implied warranties of merchantability and tness for a particular purpose Exclusive Remedies The remedies provided herein are buyer s sole and exclusive remedies Agilent Technologies shall not b...

Page 9: ...of fault A terminal marked with this symbol must be connected to ground in the manner described in the installation Operation manual and before operating the equipment Frame or chassis terminal A conn...

Page 10: ...APC 7 R is a U S registered trademark of the Bunker Ramo Corporation viii...

Page 11: ...e For Shipment 2 5 Measurement Error Correcting Function for Each Instrument 2 6 3 Operation Introduction 3 1 Operating Flow 3 1 Overview 3 2 How to set the SHORT condition and the OPEN condition 3 3...

Page 12: ...nufactured earlier than 1996 4 9 4 8 Shorting Device Position 4 10 Tables 1 1 Furnished Accessories 1 2 1 2 Option 701 Shorting Device Residual Inductance Typical 1 4 1 3 Option 010 Shorting Device Re...

Page 13: ...ly with the following LCR meters and impedance analyzers You can directly connect the 16192A to 4191A 4291B E4991A You can use the 16192A with 43961A 4395A option 010 4396A B option 010 You can use th...

Page 14: ...P N 16191 29001 1 P N 16191 29002 1 P N 16191 29003 1 P N 16191 29004 1 Case for shorting devices P N 1540 0692 1 Option 010 1 P N 16191 29005 1 P N 16191 29006 1 P N 16191 29007 1 P N 16191 29008 1 M...

Page 15: ...e not speci cations but are typical characteristics included as additional information for the operator Supplemental performance characteristics are not guaranteed Frequency Range DC to 2 GHz Applicab...

Page 16: ...ence value This method is useful for devices with values which are close to the short conditions of the measurement system The reference inductance values presented Table 1 2 and Table 1 3 were simula...

Page 17: ...d mm h mm O set mm Connector Inductance Typical P N 16191 29005 1 0 0 5 0 5 0 75 3 5 mm 0 5 nH P N 16191 29006 1 6 0 8 0 8 0 45 3 5 mm 0 4 nH P N 16191 29007 2 0 1 2 0 8 1 5 7 mm 0 9 nH P N 16191 290...

Page 18: ......

Page 19: ...so be in perfect electrical condition Verify this by carefully performing an incoming inspection to check the parallel electrode SMD test xture set for signs of physical damage and missing contents If...

Page 20: ...P N 16191 29003 1 2 422 424 5 mm P N 16191 29004 1 4 Case for shorting device P N 1540 0692 1 5 Magnifying lens34 P N 16193 60002 1 6 Tweezers4 P N 8710 2081 1 7 Insulator P N 16192 09002 1 1 Operatio...

Page 21: ...sting the Con guration to SMD Size The 16192A s applicable SMD size is 1 mm to 20 mm changing the electrodes con guration as follows Figure 2 2 Electrodes con guration and SMD Size Preparation for Use...

Page 22: ...R connector plane before connecting test xture See Table 2 2 for the applicable error correcting function After calibration disconnect any standards from the APC 7 R connector 2 Place the xture on th...

Page 23: ...als If this material is not available comparable packaging materials may be used Wrap the parallel electrode SMD test xture in heavy paper and pack in anti static plastic packing material Use su cient...

Page 24: ...lecting Test Fixture Port Extension OPEN Compensation SHORT Compensation LOAD Compensation1 4191A Calibration 0 S 0 50 Electrical Length No capability No capability No capability 4194A with 16085B No...

Page 25: ...asurement instrument s OPEN and SHORT compensation capabilities must be used 7 indicates procedures in the following pages Described in Chapter 2 The 16192A s electrical length is 11 mm typical value...

Page 26: ...e Insertion clip Tightens or loosens an electrode Shorting device Not shown in Figure 3 1 Used for short compensation If you ordered the Option 010 or 701 four types of shorting devices are supplied w...

Page 27: ...above 1 Loosen the three knobs 2 Move stage A and the two device holders to the outside 3 Adjust stage C so that the shorting device is placed on position s 4 Place the shorting device so that it con...

Page 28: ...the electrode on the stage A slightly contacts the shorting device 8 Tighten knob A while pushing the lever 9 Release the lever to hold the shorting device Setting the OPEN condition 1 Set the DUT on...

Page 29: ...How to connect the DUT 1 Push the lever and place the DUT on the electrode 2 Release the lever to hold the DUT Operation 3 5...

Page 30: ......

Page 31: ...i ed personnel only Do NOT perform any servicing other than that contained in the operating section unless you are quali ed to do so Caution When you repair the 16192A put on lint free groves to avoid...

Page 32: ...5 16191 29001 1 Shorting Device 12122 4 mm 2 6 16191 29002 1 Shorting Device 1 622 422 mm 2 7 16191 29003 1 Shorting Device 2 422 423 2 mm 2 8 16191 29004 1 Shorting Device 2 422 424 5 mm 2 9 1540 06...

Page 33: ...disassemble and assemble the Main Assembly Length Positioning Stage Assembly Removal 1 Loosen the Length Locking Knob 1 in Table 4 3 2 Remove the Length Positioning Stage Assembly 1 in Table 4 2 Width...

Page 34: ...part number and the quantity required Figure 4 2 Replaceable Parts for Main Assembly Table 4 2 Replaceable Parts for Main Assembly Reference Designator Agilent Part Number Qty Description 1 not assig...

Page 35: ...gnator Agilent Part Number Qty Description 1 16192 24006 1 Length Locking Knob 2 16192 20002 1 Length Positioning Stage Block 3 16192 23001 2 Nut 4 16192 01202 1 Contact 5 0515 2421 2 Screw M1 4L3 6 n...

Page 36: ...Replaceable Parts for Insertion Clip Assembly Reference Designator Agilent Part Number Qty Description 1 0520 0133 1 Screw 2 2190 0014 1 Washer 3 16092 40010 1 Holder 4 16092 40016 1 Holder 5 16092 40...

Page 37: ...Positioning Stage Assembly Reference Designator AgilentPart Number Qty Description 1 16192 24006 2 Width Positioning Knob 2 16192 24005 2 Device Holder Adjustment Knob 3 16192 24004 2 Holder 4 16192...

Page 38: ...Figure 4 6 Replaceable Parts for Test Stage Assembly 4 8 Service...

Page 39: ...ad M2L4 8 1250 0907 1 Contact RF Connector Caution If your 16192A test xture was manufactured earlier than 1996 and you need to replace either the plate Figure 4 7 3 or the contact assembly Figure 4 7...

Page 40: ...ng Device Position 2 Place the xture on the calibrated APC 7 R terminal of the 4291B RF Impedance Material Analyzer 3 Set the 4291B RF Impedance Material Analyzer as follows Measurement Parameter Cp p...

Page 41: ...s follows Table 4 8 Short Impedance Value Guideline Parameter Frequency Guideline absolute value Ls 100 MHz 4 5 nH 6 1 5 nH Ls 1 GHz 4 5 nH 6 1 5 nH Short Impedance Repeatability Check 1 After the Sho...