Operation
Cleaning Connectors

Cleaning Connectors

Clean connectors are essential for ensuring the integrity of RF coaxial 

connections. Use the following procedure to clean the connectors.

1. Use compressed air or nitrogen

Use compressed air (or nitrogen) to loosen particles on the connector 

mating plane surfaces. Clean air can’t damage a connector, or leave 
particles or residues behind.

You can use any source of clean, dry low-pressure compressed air or 

nitrogen that has an effective oil-vapor filter liquid condensation trap 

placed just before the outlet hose. Ground the hose nozzle to prevent 

electrostatic discharge, and set the air pressure to less than 414 kPa to 

control the velocity of the air stream. High-Velocity streams of compressed 

air can cause electrostatic effects when directed into a connector.

2. Clean the connector threads

Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to 

remove any dirt or stubborn contaminants on a connector that can’t be 

removed with compressed air or nitrogen.

a. Apply a small amount of isopropyl alcohol to the lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle 

stream of clean, low-pressure compressed air or nitrogen.

3. Clean the mating plane surfaces

Apply a small amount of isopropyl alcohol to a new swab and clean the 

center and outer conductor mating plane surfaces.

4. Dry the connector

After cleaning, blow the connector dry with a gentle stream of clean 

compressed air or nitrogen. Always completely dry a connector before you 

reassemble or use it.

5. Reinspect 

Inspect the connector again to be sure that no particles or residue.

Always use protective eyewear when using compressed air or nitrogen.

Summary of Contents for 16190B

Page 1: ...Operation and Service Manual Keysight 16190B Performance Test Kit...

Page 2: ...puter software under the same terms by which the software is customarily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial lic...

Page 3: ...8 Electrical Characteristic 8 Pin Depth 8 Operating Conditions 8 Storage Conditions 8 Carrying Case Dimensions 9 2 Operation Operating Precautions 11 Handling and Storage 11 Connection Techniques 11 C...

Page 4: ...4 Keysight 16190B Operation and Service Manual Contents...

Page 5: ...lyzers or impedance meters that have a 7 mm measurement terminal Kit Contents The 16190B contents are listed in Table 1 1 Table 1 1 Contents of the 16190B Description Keysight Part No Qty Airline 50 7...

Page 6: ...s and the performance test kit has been checked mechanically and electrically The shipping contents should consist of the items shown in Table 1 1 If the shipment is incomplete or if there is mechanic...

Page 7: ...efer to the unit by model number and full serial number Non Keysight Packing Materials The following general instructions should be used for re packing with commercially available packing materials 1...

Page 8: ...ble 1 4 Storage Conditions The kit must be stored under the ambient environmental conditions in Table 1 5 Table 1 2 Electrical Characteristics OPEN Termination 0 63 pF 5 at 1 MHz 50 Termination 50 0 5...

Page 9: ...16190B Performance Test Kit 9 General Information Characteristics Carrying Case Dimensions The carrying case dimensions are 350 mm W x 100 mm H x 270 mm D Altitude 0 m to 15000 m Table 1 5 Storage Co...

Page 10: ...10 Keysight 16190B Performance Test Kit General Information Characteristics 1...

Page 11: ...its tube replace the airline outer conductor and the termination s protection caps and store the devices in the carrying case Connection Techniques Extreme care should be taken when making connections...

Page 12: ...on 2 collets PN 85050 20001 If the inspection shows that the airline center conductor tip needs to be replaced replace the tip PN 1250 2006 Do not use damaged connectors Document Microwave Connector C...

Page 13: ...cause electrostatic effects when directed into a connector 2 Clean the connector threads Use a lint free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contami...

Page 14: ...eeve 2 While pressing terminations from the top as shown inFigure 2 2 firmly connect it to the 7 mm terminal clockwise viewed from the top Figure 2 2 Termination connection When making connections wit...

Page 15: ...Fully retract the threads on the 7mm connector Then insert the marked side tip of the airline center conductor Figure 2 3 Airline connection to 7mm connector b Gently cover the airline center conducto...

Page 16: ...performance test This section provides the diskette format information that is necessary to use the diskette Table 2 1 outlines the calibration data diskette format Table 2 1 Diskette Format Outline...

Page 17: ...satisfy all of these calibration uncertainties For complete information on calibration contact your nearest Keysight Technologies office Table 3 1 Impedance Magnitude Z Admittance Magnitude Y Calibrat...

Page 18: ...requency MHz 50 Airline with OPEN Airline with SHORT 1 1 0 mrad 1 6 mrad 1 7 mrad 10 1 0 mrad 1 6 mrad 1 7 mrad 100 1 0 mrad 1 6 mrad 1 8 mrad 200 2 0 mrad 2 0 mrad 2 0 mrad 300 2 0 mrad 2 0 mrad 2 0...

Page 19: ...n on repair contact your nearest Keysight Technologies office The cap for Open Termination Short Termination 50 Termination and Airline 50 7mm can be ordered with the part number as below 2400 8 0 mra...

Page 20: ...20 Keysight 16190B Performance Test Kit Service Repair 3...

Page 21: ...This information is subject to change without notice Keysight Technologies 2000 2017 Edition 4 October 2017 16190 90020 16190 90020 www keysight com...

Reviews: