Special Supplement

8

Keysight 11683AZ Option H01 Operating and Service Manual

Description

The 11683AZ Option H01 is a standard 11683A range calibrator that has been modified so that it 

will operate with either the Internal DC-Reference source, or an external programmable 

DC-Reference source.

Specifications

External DC Reference is the same as the Internal Reference.

Operation

The 11683AZ Option H01 range calibrator allows a user to select either an internal or an external 

DC-Reference source. With the 

REF SELECT

 switch in the 

INT

 position, the range calibrator is 

operating manually with the Internal DC-Reference source. With the 

REF SELECT

 switch in the 

EXT

 

position, the range calibrator can be programmed remotely with a programmable DC source. In 

this mode of operation, the 

REF SELECT

 switch disconnects the internal DC source, and connects 

the external DC source to the sampling gate assembly (A3).
In all other respects, the 11683AZ Option H01 operates the same as the standard 11683A.
The typical input voltages corresponding to the different range settings are:

The above calculated voltages for range 100 mW to 300 

μ

W are based on the table shown in the 

11683A Operating and Service Manual (11683-90014). The remaining voltages are based in the 

voltage divider circuit (A1), and the input resistance (A1R1, A3A1U1R1 and A3R1), of the sampling 

gate assembly shown in the 11683A Operating and Service Manual (11683-90014).

NOTE

Note that the 11683AZ Option H01 is not equivalent to the standard 11683A. When ordering this 

product, you will need to order 11683AZ Option H01 instead of 11683A.

Range

Volts

100 mW

15.800 V

30 mW

4.7121 V

10 mW

1.4641V

1 mW

145.00 mV

3 mW

458.00 mV

300 

μ

W

45.837 mV

100 

μ

W

14.494 mV

30 

μ

W

4.5832 mV

10 

μ

W

1.4501 mV

μ

W

458.69 

μ

V

Summary of Contents for 11683AZ Option H01

Page 1: ...Keysight 11683AZ Option H01 Power Meter Range Calibrator Operating and Service Manual ...

Page 2: ... these rights customarily provided to the public to use modify reproduce release perform dis play or disclose commercial computer soft ware or commercial computer software documentation No additional government requirements beyond those set forth in the EULA shall apply except to the extent that those terms rights or licenses are explicitly required from all providers of commercial computer softwa...

Page 3: ...sight 11683AZ Option H01 Operating and Service Manual 3 Table of Contents 1 Special Supplement Description 8 Specifications 8 Operation 8 Performance Test 9 Internal DC reference 9 External DC reference 9 ...

Page 4: ...4 Keysight 11683AZ Option H01 Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 5: ...Keysight 11683AZ Option H01 Operating and Service Manual 5 List of Tables Table 1 1 Digital voltmeter reading OHM 10 ...

Page 6: ...6 Keysight 11683AZ Option H01 Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 7: ...Keysight 11683AZ Option H01 Power Meter Range Calibrator Operating and Service Manual 7 1 Special Supplement Description 8 Specifications 8 Operation 8 Performance Test 9 ...

Page 8: ...ion the REF SELECT switch disconnects the internal DC source and connects the external DC source to the sampling gate assembly A3 In all other respects the 11683AZ Option H01 operates the same as the standard 11683A The typical input voltages corresponding to the different range settings are The above calculated voltages for range 100 mW to 300 μW are based on the table shown in the 11683A Operati...

Page 9: ...ed equipment for performing these tests are a digital voltmeter with 5 digits resolution a jumper 18 awg or heavier wire and a banana connector to the BNC 4 wire cable Procedure Set the range as follows 1 Set the digital voltmeter controls for automatic DC voltage measurements 2 Connect one end of the 4 wire cable to the digital voltmeter and the other end to the DC REFERENCE OUTPUT on the rear pa...

Page 10: ... 300 mW to 3 mW to 5 digit resolution and record the reading on the table below Verify that each reading falls within the limits shown 10 Disconnect the 4 wire cable and jumper from the 11683AZ H01 This concludes the external DC reference circuit performance test If any of the voltage or resistance readings are incorrect refer to the troubleshooting information in the 11683A Operating and Service ...