Using Probe Heads
2
1130B-Series Probes User’s Guide
73
12. E2678B Differential Socketed Head with Header Adapter
This probe configuration can be used to connect to 25 mil square pin headers with
100 mil spacing such as those used in USB testing. The header adapter is
recommended for use with the 1130B and 1131B InfiniiMax probes.
All of the specifications and characteristics of the header adapter are the same as
those for the socketed differential probe head except for the input capacitance
shown in
To adapt the 01130-21303 damped wire accessory from solder-in to plug-on,
solder the tip into the 01131-85201 square pin socket and then slip the
01131-41101 heat-shrink sleeve over the solder joint and heat the heat-shrink
tubing with a heat gun. This allows the damped wire accessories to be used to
plug onto 25 mil square pins.
Figure 44
01130-63201 Header Adapter Dimensions
NOTE
If the header adapter is used with higher bandwidth probe amplifiers such as the 1132B (5
GHz) or the 1134B (7GHz), the rise time of the input signal should be slower than ~150 ps
(10% to 90%) to limit the effects of resonances in the adapter.
NOTE
The E2678B is a direct replacement for the E2678A that provides a more robust connection
between the E2678B head's socket and printed circuit board. The E2678B’s performance and
fit are identical to that of the E2678A.
Table 25
Characteristic Capacitance
Identification
Capacitance
Description
Cm
0.43 pF
Model for input C is Cm between the tips and Cg to ground each tip
Cg
0.54 pF
Cdiff
0.70 pF
Differential mode capacitance is Cm + Cg/2
Cse
0.97 pF
Single-ended mode capacitance is Cm + Cg
Summary of Contents for 1130B Series
Page 1: ...Keysight 1130B Series Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1130B Series Probes User s Guide ...
Page 59: ...Using Probe Heads 2 1130B Series Probes User s Guide 59 Figure 33 N5380B Schematic ...
Page 78: ...78 1130B Series Probes User s Guide 2 Using Probe Heads ...
Page 86: ...86 1130B Series Probes User s Guide 3 Calibrating Probes Figure 53 Overlapping Waveforms ...
Page 90: ...90 1130B Series Probes User s Guide 3 Calibrating Probes ...
Page 98: ...98 1130B Series Probes User s Guide 4 Characteristics and Specifications ...
Page 226: ...226 1130B Series Probes User s Guide 10 SPICE Models ...
Page 228: ...228 1130B Series Probes User s Guide 10 SPICE Models ...