
24
1130B-Series Probes User’s Guide
1
Getting Started
N2880A InfiniiMax In-Line Attenuator Kit
The in-line attenuators are an accessory for probes. The maximum input range of
the 1130B probes are 5 V
p-p
. If you need to measure larger signals, the probe’s
design allows you to add the N2880A InfiniiMax in-line attenuators between the
probe head and probe amplifier to increase the maximum input range (as listed in
on page 25). Additionally, these attenuators enable you to increase the
offset range of the probe as specified in
N2880A In-Line Attenuators, the bandwidth and rise time of your probing system
is not affected. There is, however, a trade-off in noise (refer to
) and in the
accuracy of DC offset relative to the input.
The N2880A provides a pair of 6 dB, 12 dB, and 20 dB attenuators. The
attenuators come as matched pairs and should only be used with each other. Each
attenuator has a serial number. The pair of matching attenuators in each set will
have the same four digit numeric prefix and will differ by the last letter (one
attenuator in the matched pair will be labeled A and the other will be labeled B).
Figure 13
Placement of Attenuators Between Probe Amplifier and Head
CAUTION
Prevent abrasion and tears in the cable’s jacket, do not rest the extension cables on
any metal objects or objects with sharp edges.
CAUTION
Do not kink the cables. The cables are designed to be flexible, but are not designed to
be bent sharply.
NOTE
Keep your extreme temperature testing probes separate from the probes they use under
milder conditions. This is because cycling probe heads through extreme temperature ranges
has a marked affect on their lifetimes as listed in
. Only the lifetime of the probe head
is affected by temperature cycling. The extension cables and probe amplifier should not need
to be replaced with extended temperature cycling.
NOTE
Discoloration or texture changes are possible with the extension cables. These changes do
not, however, affect the performance or the quality of a measurement.
Summary of Contents for 1130B Series
Page 1: ...Keysight 1130B Series Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1130B Series Probes User s Guide ...
Page 59: ...Using Probe Heads 2 1130B Series Probes User s Guide 59 Figure 33 N5380B Schematic ...
Page 78: ...78 1130B Series Probes User s Guide 2 Using Probe Heads ...
Page 86: ...86 1130B Series Probes User s Guide 3 Calibrating Probes Figure 53 Overlapping Waveforms ...
Page 90: ...90 1130B Series Probes User s Guide 3 Calibrating Probes ...
Page 98: ...98 1130B Series Probes User s Guide 4 Characteristics and Specifications ...
Page 226: ...226 1130B Series Probes User s Guide 10 SPICE Models ...
Page 228: ...228 1130B Series Probes User s Guide 10 SPICE Models ...