6
ALL testing functions have the following
Touch Pad
Gives an alert, when touching the Touch Pad, while the PE
terminal is connected to Line by mistake.
Auto power off Automatically switches the instrument off after a period of
approximately
10
minutes. The Auto power off mode can
only be cancelled by switching the instrument on again.
5
.
1
Measurement Specification
Continuity
Open Circuit
Voltage (DC)
Short Circuit
Current
Range
Accuracy
5
V
±
20
%
Greater than
200
mA
20
/
200
/
2000
Ω
Auto - Ranging
0
~
0
.
19
Ω
±
0
.
1
Ω
0
.
2
~
2000
Ω
±
(
2
%rdg+
8
dgt)
2
Ω
Buzzer : Buzzer sounds when measured resistance is
2
Ω
or less.
2
Ω
Buzzer Accuracy :
2
Ω
±
0
.
4
Ω
Insulation Resistance
Open Circuit
Voltage (DC)
Rated Current
Range
Accuracy
250
V+
40
% -
0
%
1
mA or greater
@
250
k
Ω
20
/
200
M
Ω
Auto - Ranging
0
~
19
.
99
M
Ω
:
±
(
2
%rdg+
6
dgt)
20
~
200
M
Ω
:
±
(
5
%rdg+
6
dgt)
500
V+
30
% -
0
%
1
mA or greater
@
500
k
Ω
20
/
200
/
2000
M
Ω
Auto - Ranging
0
~
199
.
9
M
Ω
:
±
(
2
%rdg+
6
dgt)
200
~
2000
M
Ω
:
±
(
5
%rdg+
6
dgt)
1000
V+
20
% -
0
%
1
mA or greater
@
1
M
Ω
20
/
200
/
2000
M
Ω
Auto - Ranging
5
Specification