20
Normal from 199.0 to 201.0
(3) After counting down, "0.00g" will be displayed for
a few seconds, and then the decimal point starts
blinking.
When the internal weighing scale displays "0.0g" as
the zero point, put the 200g standard weight gently
on the sample cup covered on the inner cylinder.
(4) The weight of the 200g standard weight is displayed.
The display example in the drawing on the right is
200.0g.
When the displayed value falls within the range from
199.0 through 201.0g, the result is normal. If the value
is above or below the range, adjustment is required.
(5) Press the
key to terminate.
This chapter, "7-1. 200g standard weight (for checking the accuracy of weighing scale)", is described
for measuring volume weight, but is also useful for inspection operations using moisture content
measurement. This procedure can be used as standard operating protocol before operation.