![KERN PBJ-N Service Manual Download Page 53](http://html1.mh-extra.com/html/kern/pbj-n/pbj-n_service-manual_1957049053.webp)
53
PBJ_N-SH-e-1810.docx
-
If the error is large, perform PCAL (calibration of the built-in weight).
For information, refer to the section on calibration of the built-in weight in the instruction
manual.
21.4 Linearity
-
Select menu “ E-CAL “ and calibration the span using the calibration weight.
-
Load the weight as shown in “ table 1 “ in turn and finally unload and take a reading.
-
Obtain the average error of each measurement. ( error ) = ( reading values ) - ( actual
weight ) - ( average of unloaded reading values )
-
Operation is normal if the error is less than the specification value.
-
If the error is large, perform Writing in regular linear coefficients
21.5 Tilt error
-
Load the weight as shown in “ table 1 “ and press the -->O/T<--
-
Insert a 1mm thick board under the two feet on the front side of the balance and take a
B ) ( error ) = ( B ) - ( A )
-
If the error is large, perform Adjustment of tilt errors.
TYPE
PBJ 620-3NM
PBJ 4200-2NM
PBJ 6200-2NM
PBJ 8200-1NM
CAPACITY
[g]
620
4200
6200
8200
Readability
[mg]
1
10
10
100
Repeatability(Rx,Ry)
weight
[mg]
2
20
20
200
[g]
300
2000
3000
4000
Corner load
error
weight
[mg]
±5
±30
±50
±200
[g]
200
1000
2000
2000
Linearity
weight
[mg]
±2
±20
±20
±100
[g]
100
200
400
600
620
1000
2000
3000
4000
4200
1000
2000
4000
6000
6200
2000
4000
6000
8000
8200
Built-in
Calibration(*1)
weight
[mg]
±5
±30
±50
±200
[g]
600
4000
6000
8000
Tilt error
weight
[mg]
±30
±300
±300
±3000
[g]
250
4000
4000
4000
Table 1
Adjustment Specification
Summary of Contents for PBJ-N
Page 4: ...PBJ_N SH e 1810 docx 4 ...
Page 40: ...PBJ_N SH e 1810 docx 40 Fig 7 Fig 8 ...
Page 47: ...47 PBJ_N SH e 1810 docx Fig 22 Fig 23 ...
Page 55: ...55 PBJ_N SH e 1810 docx 23 Drawing Fig 31 ...
Page 56: ...PBJ_N SH e 1810 docx 56 ...
Page 57: ...57 PBJ_N SH e 1810 docx Fig 33 ...