13.4 Error messages
Error message
Description
Possible causes
Zeroing range exceeded
due to switching-on balance
or pressing
(normally
4% max)
•
Object on the weighing
plate
•
Overload when zeroing
•
Improper adjustment
•
Damaged weighing cell
•
Damaged electronics
Keyboard error
•
Improper operation of the
balance
Value outside the A/D
changer range
•
Damaged weighing cell
•
Damaged electronics
Unable to initialise zero
point
•
Measuring cell defective
/ overloaded
•
Object on weighing pan
/ contact
•
Transport safety device
has not been removed
•
Main board defective
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
ERR4
0
ERR5
ERR6