B.16 SOURce:LIST:RTIMe COMMAND
SOUR:LIST:RTIM
Syntax:
Short Form: SOUR:LIST:RTIM <NRf> Long Form: SOURce:LIST:RTIMe <NRf>
Description:
Sets the output ramp-up time.
Argument:
<NRf> is a value in the range from 0.0 to 9.9 seconds.
Example:
SOURce:LIST:RTIMe 3.0 programs the output ramp-up time to 3.0 seconds.
B.17 SOURce:LIST:RTIMe?
QUERY
SOUR:LIST:RTIM?
Syntax:
Short Form: SOUR:LIST:RTIM?
Long Form: SOURce:LIST:RTIMe?
Description:
Indicates (in seconds) how long output voltage takes to reach programmed value.
Return Format: <NRf>
Example:
SOURce:LIST:RTIMe? returns 3.0 to indicate ramp-up time is 3.0 seconds.
B.18 SOURce:MEMory:CLS COMMAND
SOUR:MEM:CLS
Syntax:
Short Form: SOUR:MEM:CLS Long
Form:
SOURce:MEMory:CLS
Description:
Clears all memory locations used for storing and recalling power supply settings.
Argument:
None
Example:
SOURce:MEMory:CLS clears all memory locations.
B.19 SOURce:MEMory:CURRent:X COMMAND
SOUR:MEM:CURR:X
Syntax:
Short Form: SOUR:MEM:CURR:X <NRf> Long Form: SOURce:MEMory:CURRent:X <NRf>
Description:
Sets the memory location defined by X to the current value <NRf>.
Argument:
<NRf> is a value in the range from 0 to rated current of unit.
Example:
SOURce:MEMory:CURRent:5 11.6 sets memory location 5 to 11.6A.
B.20 SOURce:MEMory:CURRent:X?
QUERY
SOUR:MEM:CURR:X?
Syntax:
Short Form: SOUR:MEM:CURR:X? Long Form: SOURce:MEMory:CURRent:X?
Description:
Returns current value stored in memory location X (in Amperes).
Return Format: <NRf>
Example:
SOURce:MEMory:CURRent:5? returns 11.6 (Voltage stored in memory location 3 is 11.6A).
B.21 SOURce:MEMory:LIST:X?
QUERY
SOUR:MEM:LIST:X?
Syntax:
Short Form: SOUR:MEM:LIST:X?
Long Form: SOURce:MEMory:LIST:X?
Description:
For the memory location defined by X, returns voltage (Volts) and current (Amperes) in sci-
entific notation.
Return Format: <NRf>
Example:
SOURce:MEMory:LIST:3? returns 5.02, 2.50000E-00 (Voltage/Current stored in memory
location 3 is 50V, 2.5A).
B.22 SOURce:MEMory:RECall:X COMMAND
SOUR:MEM:REC:X
Syntax:
Short Form: SOUR:MEM:REC:X
Long Form: SOURce:MEMory:RECall:X
Description:
Restores power supply settings to the voltage and current values stored in memory loca-
tion defined by X.
If the output is off, the recalled voltage and current set values are shown on the
front panel Voltage and Current displays. If the output is on, the output changes to the recalled
voltage and current values which are also shown on the front panel Voltage and Current displays.
Argument:
X is a number from 0 to 15 representing one of 16 memory locations
Example:
SOURce:MEMory:RECall:3 sets power supply setpoints to 50V and 2.5A (values previously
stored via either front panel or SOURce:MEMory:VOLTage:X and SOURce:MEMory:CURRent:X
commands). If output is on, output is changed to 50V, 2.5A. If output is off, set values of 50V and
2.5A are displayed on the front panel.