BHK-MG (OPR) 032816
3-17
3.3.2
DCL CONTROL
The device clear (DCL) and selected device clear can be set to operate in two modes. In the
“output = 0” mode, when the device clear is received the output is disabled; when the output is
enabled, the output of the power supply is set to zero volts and current is set to the power-up
and reset current value (see PAR. 3.2.8.1). In the “output unchanged” mode sending DCL or
selected DCL disables the output, but has no effect on output voltage and current as required by
IEEE specification 488.2 so that when the output is enabled, the settings in effect when DCL
was issued are restored. Note that the DCL setting also affects the operation of RESET (see
PAR. 3.2.8). The factory default value is “output=0” mode.
To change the DCL mode the unit must be in command entry mode
(:_:_:)
, then depress the
MENU
key until the top line of the display shows the current setting: either
DCL OUTP = 0
or
DCL = OUTP UNCHNG
. Depress
1
for “output unchanged” mode where the output is unaffected
by DCL; depress
0
for “output = 0V” mode where DCL sets the output to 0 volts and current is
set to the power-up and reset current value (see PAR. 3.2.8.1).
3.3.3
CHANGING THE GPIB ADDRESS
When the power supply is in command entry status
(:_:_:)
, press
MENU
key until the top line
of the LCD indicates the current GPIB address. Enter new value and press
ENTER
to accept
new setting, or
CLEAR
to exit without changing setting. The GPIB address default value is set
to 6.
3.3.4
EXTERNAL TRIGGER
The External Trigger Port can be used to restore the power supply output to settings previously
defined using SCPI commands. This can be done by applying a temporary short between pin 4
and pin 1 of the External Trigger Port, (or if pin 4 is at TTL logic 1, applying a temporary logic 0
to pin 4 referenced to pin 1).
NOTE: Turn off power to the unit or disable the output before inserting the plug in the External
Trigger port to avoid unintentionally initiating shutdown or trigger.
The following remote SCPI commands must be sent before the external trigger at pin 4 acti-
vated:
•
The values to be set upon receiving the external trigger at pin 4 must be programmed
using VOLT:TRIG (PAR B.58) and CURR:TRIG (PAR B.48) commands.
TABLE 3-8. IEEE 488 (GPIB) BUS DATA MODE MESSAGES
MNEMONIC
MESSAGE DESCRIPTION
COMMENTS
DAB
Data Byte
Received or Sent
END
End
Received or Sent
EOS
End of String
Received or Sent
RQS
Request Service
Sent
STB
Status Byte
Sent
Summary of Contents for BHK 1000-0.2MG
Page 2: ......
Page 6: ......
Page 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Page 18: ...xii BHK MG032816 ...
Page 46: ......
Page 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Page 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Page 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Page 116: ......
Page 154: ......