Operating manual
6
Do 1 source ...... ready-signal
Do 1 function
level/signal
Unit .................. Celsius
Do 1 delay time ...... 0.00 s
Terminal assigned to ...... USB
Do 1 hold time ...... 0.00 s
Autoprint .................. on
Do 2 source ............... off
Print cycle time ...... 0.1 s
Protocol address .......... 001
Display ................ active
Key lock .................. off
Status LED ... assigned to Do 1
------------------------------------------------------------------
>
The parameters of the measured value recording quotient are listed at
the top left. The I/O configuration is in the right column. The general set-
tings are on the bottom left.
A detailed description of the individual menu items can be found in the
pyrometer operating instructions.
6.2.3 Description of the Mixed Mode submenu
In the sub menu „Mixed Mode“ you can adjust all relevant parameters In
the sub menu „Mixed Mode“ you can adjust all relevant parameters:
Activation/ deactivation of the temperature calculation
Measuring range of the Mixed Mode
Beginning off the transition zone
Selection of the first source (100 % temperature below the transi-
tion zone)
Selection of the second source (100 % temperature above the
transition zone)
Selection of the controlling source (temperature control for transi-
tion zone)
>M
------------------------------------------------------------------
Submenu MIXED-MODE
------------------------------------------------------------------
MIX function ............... on
MIX range ... 500.0 - 2400.0 C
MIX transition 680.0 - 700.0 C
1st source .......... lambda 1
2nd source .......... quotient
ctrl source .......... lambda 1
M: Function Mixed-Mode
A: Set range begin
B: Set range end
C: Set transition begin
D: Set transition end
E: Set 1st source
F: Set 2nd source
G: Set ctrl source
X: Show mixed temperature
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>MIXED MODE >