Technical Data PT 128 AF 10
Measuring range:
75 ... 650 °C
Sensors:
photo diode
Spectral sensitivity:
1.8
– 2.4 µm
Response time t
98
:
≤ 200 ms (T > 75 °C)
≤ 50 ms (T > 100 °C)
≤ 15 ms (T > 125 °C)
≤ 2 ms (T> 200 °C)
Measuring uncertainty:
0.75 % of range and but at
least 3 K
(ε =1.0 and T
A
=23
°C)
Repeatability:
1 K
Temperature coefficient with
reference to 23 °C:
0.25 K / K (for T < 500 °C)
0.05 % / K (for T ≥ 500 °C)
of measured value