Models 2510 and 2510-AT
Performance Verification
1-12
4.
Configure the Model 2510 for thermistor sensor type, resistance range, and 4-wire
sense mode as follows:
•
Press CONFIG then T. The instrument will display the following:
CONFIG TEMPERATURE
PROTECTION SENSOR-TYPE UNITS PID
•
Select SENSOR-TYPE, and then press ENTER. The following will be displayed:
SENSOR TYPE
THERMISTOR RTD I-SS V-SS
•
Select THERMISTOR, and then press ENTER. The following will be displayed:
THERMISTOR SETTINGS
RANGE A B C I-SRC
©
•
Select RANGE, then press ENTER. The unit displays:
THERMISTOR RANGE
100 1K 10K 100K
•
Select 100, then press ENTER.
•
From the THERMISTOR SETTINGS menu, choose SENSE-MODE, then press
ENTER.
•
From the SENSE MODE menu, choose 4-WIRE, then press ENTER.
•
Press EXIT as required to return to normal display.
5.
Repeatedly press the DISPLAY TOGGLE key until the sensor resistance (RT) is dis-
played on the lower display line on the right. For example: RT: +099.98
Ω
.
6.
Verify that the sensor reading for the 100
Ω
resistance in Table 1-3 is within stated
limits.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
120
LINE RATING
100-240VAC
50, 60 HZ
90VA MAX
LINE FUSE
SLOWBLOW
2.5A, 250V
!
!
CAT
I
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
ENABLE-DIG I/O
RS-232
TRIGGER
LINK
OUTPUT
INPUT
F+ S+ S- F- F+ F- S+ S-
ISOLATION FROM EARTH: 30V MAX.
Model 2510
100
Ω
, 1k
Ω
, 10k
Ω
,
or 100k
Ω
Resistor
INPUT S-
INPUT F-
INPUT S+
INPUT F+
Connect sense (S) leads as close
as possible to resistor body.
Figure 1-4
Connections for sensor
resistance accuracy
verification
2510-902-01 (E - Mar 2010)(Service).book Page 12 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...