Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. B / May 2013
2-55
Multiple SMU connections
Carefully consider and configure the appropriate output-off state, source, and compliance limits
before connecting the Series 2600B to a device that can deliver energy (for example, other voltage
sources, batteries, capacitors, solar cells, or other Series 2600B instruments). Configure
recommended instrument settings before making connections to the device. Failure to consider the
output-off state, source, and compliance limits may result in damage to the instrument or to the
device under test (DUT).
The following figures show how to use the SMUs of two Series 2600B instruments to test a 3-terminal
device, such as an N-channel JFET (see
TSP advanced features
(on page 6-53) for information on
using multiple Series 2600B instruments). A typical application is for a Series 2600B to source a
range of gate voltages, while the another Series 2600B sources voltage to the drain of the device and
measures current at each gate voltage.
Figure 22: Two SMUs connected to a 3-terminal device (local sensing)
Summary of Contents for System SourceMeter 2601B
Page 841: ......