Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-237
Example
smua.trigger.endpulse.action =
smua.SOURCE_IDLE
smua.trigger.endpulse.stimulus =
trigger.timer[1].EVENT_ID
Configure the end pulse action to achieve a
pulse and configure trigger timer 1 to control
the end of pulse.
Also see
Triggering
(on page 3-32)
smuX.trigger.endpulse.set()
This function sets the end pulse event detector to the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu
X
.trigger.endpulse.set()
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.set()
applies to SMU channel A)
Details
This function sets the end pulse event detector to the detected state.
The SMU automatically clears all the event detectors when the
smu
X
.trigger.initiate()
function is
executed. Therefore,
smu
X
.trigger.endpulse.set()
should be called after the sweep is initiated. If the
event detectors are configured to clear automatically because the
smu
X
.trigger.autoclear
attribute is set to
smu
X
.ENABLE
, make sure that
smu
X
.trigger.endpulse.set()
is issued after the SMU has entered the
trigger layer.
Also see
smuX.trigger.autoclear
(on page 7-235)
smuX.trigger.initiate()
(on page 7-240)
Triggering
(on page 3-32)
smuX.trigger.endpulse.stimulus
This attribute defines which event will cause the end pulse event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0
Usage
eventID
= smu
X
.trigger.endpulse.stimulus
smu
X
.trigger.endpulse.stimulus =
eventID
eventID
Set to the event that triggers the end pulse action of the source
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.stimulus
applies to SMU channel A)