Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-193
Example
-- Short SENSE LO and LO terminals
-- Short SENSE HI and HI terminals
-- Allow readings to settle, then get measurements
r0_hi, r0_lo = smua.contact.r()
-- Connect 50 OHM resistor between SENSE LO and LO
-- Connect 50 OHM resistor between SENSE HI and HI
-- Allow readings to settle, then get measurements
r50_hi, r50_lo = smua.contact.r()
smua.contact.calibratelo(r0_lo, Z_actual_lo, r50_lo,
50_ohm_actual_lo)
smua.contact.calibratehi(r0_hi, Z_actual_hi, r50_hi,
50_ohm_actual_hi)
Performs contact
check on SMU
channel A.
Install and measure
two resisters.
The user sends
contact check LO
calibration
command.
The user sends
contact check HI
calibration
command.
Also see
Adjustment
(on page B-18)
smuX.cal.restore()
(on page 7-188)
smuX.cal.save()
(on page 7-189)
smuX.cal.unlock()
(on page 7-190)
smuX.contact.calibratehi()
smuX.contact.check()
This function determines if contact resistance is lower than the threshold. This function is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu
X
.contact.check()
X
Source-measure unit (SMU) channel (for example,
smua.contact.check()
applies to SMU channel A)