Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-82
2600BS-901-01 Rev. B / May 2013
Example
key = display.waitkey()
print(key)
Pause script execution until the operator presses
a key or the navigation wheel
, or rotates the
navigation wheel.
If the output is:
8.01
It indicates that the MEAS(A) key was pressed.
Also see
Capturing key-press codes
(on page 3-81)
display.getlastkey()
(on page 7-63)
display.sendkey()
display.settext()
(on page 7-76)
errorqueue.clear()
This function clears all entries out of the error queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
errorqueue.clear()
Details
See the
Error queue
(on page E-3) topic for additional information about the error queue.
Also see
Error queue
(on page E-3)
errorqueue.count
(on page 7-83)
errorqueue.next()
(on page 7-83)
Summary of Contents for System SourceMeter 2601B