Section 4: Theory of operation
Series 2600B System SourceMeter® Instrument Reference Manual
4-4
2600BS-901-01 Rev. B / May 2013
When attempting to determine the maximum duty cycle, where the off state will be 0 V or 0 A:
I
B
is 0
I
P
and V
P
are the voltage and current levels when the instrument is on
Model 2601B/2602B/2604B maximum duty cycle equation constants
Constant
100 mV range 1 V range
6 V range
40 V range
V
OA
18
18
18
55
Model 2611B/2612B/2614B/2634B/2635B/2636B maximum duty cycle equation
constants
Constant
200 mV range 2 V range
20 V range
200 V range
V
OA
40
40
40
220
Operating boundaries
Source or sink
Depending on how it is programmed and what is connected to the output (load or source), the
instrument can operate in any of the four quadrants. The four quadrants of operation are shown in the
continuous operating boundaries figures. When operating in the first (I) or third (III) quadrant, the
instrument is operating as a source (V and I have the same polarity). As a source, the instrument is
delivering power to a load.
Figure 91: Four quadrants of operation
When operating in the second (II) or fourth (IV) quadrant, the instrument is operating as a sink (V and
I have opposite polarity). As a sink, it is dissipating power rather than sourcing it. An external source
or an energy storage device, such as a capacitor or battery, can force operation in the sink region.