Command Reference B-9
Error 505
— Will occur when a :CAL:PROT:SOUR command generates an invalid source off-
set calibration constant.
Error 506
— Will occur when a :CAL:PROT:SOUR command generates an invalid source gain
calibration constant.
Error 50
7 — Will occur when a :CAL:PROT:SENS command generates an invalid measure-
ment offset calibration constant.
Error 508
— Will occur when a :CAL:PROT:SENS command generates an invalid measure-
ment gain calibration constant.
Table B-4
Calibration errors
Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it
is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query to
read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an er-
ror occurs, send the following command: *SRE 4. This command will enable SRQ when the
EAV bit is set. You can then read the status byte and error queue as outlined above to check for
errors and to determine the exact nature of the error.
Error
number
Error message
+500
"Date of calibration not set"
+501
"Next date of calibration not set"
+502
"Calibration data invalid"
+503
"DAC calibration overflow"
+504
"DAC calibration underflow"
+505
"Source offset data invalid"
+506
"Source gain data invalid"
+507
"Measurement offset data invalid"
+508
"Measurement gain data invalid"
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...